What to do if ConMed Linvatec Hall PowerPro PRO6150 Medical Equipment handpiece lacks power?
- EEvan NielsenOct 31, 2025
If you're experiencing a lack of handpiece power, there are several potential causes to investigate: * A regulator malfunction could be the issue. Try running the handpiece on another regulator to determine if the problem lies with the handpiece itself or the regulator. Replace the faulty equipment. * Incorrect operating pressure can also lead to power issues. Ensure the pressure is set to the recommended operating level. * Check all hose connections to confirm they are fully and properly seated in both the regulator and the handpiece. * Restrictions in the hose can impede performance. Remove any obstructions. * Do not start the procedure if the tank pressure is below 500 psi. Replace the tank. * Ensure the tank valve is completely open. * Make sure the mode lever is in the ...