13
Pause the alarm sound indication
Open the alarm sound indication
Close the pulse sound indication
Open the pulse sound indication
menu button/power button/function button
Type BF
SN Serial number
1. the finger clip falls off ( no finger inserted)
2. Probe error
3. Signal inadequacy indicator
IP22 International Protection
WEEE (2002/96/EC)
10 Function Specification
Information Display Mode
The Pulse Oxygen Saturation(SpO
2
) 2-digit digital OLED display
Pulse Rate(PR) 3-digit digital OLED display
Pulse Intensity (bar-graph) bar-graph OLED display
SpO
2
Parameter Specification
Measuring range 0% ~ 100%, (the resolution is 1%).
Accuracy 70% ~ 100%:±2% ,Below 70% unspecified.
Average value
Calculate the Average value in every 4 measure value.
The deviation between average value and true value does not
exceed 1%.
Pulse Parameter Specification
Measuring range 30bpm ~ 250bpm, (the resolution is 1bpm)
Accuracy ±2bpm or±2% (select larger)