1 Safety
1.1 Instructions for Safe Operations
Check the main unit and all accessories periodically to
make sure that there is no visible damage that may affect
patient’s safety and monitoring performance about cables
and transducers. It is recommended that the device should
be inspected once a week at least. When there is obvious
damage, stop using the device. In addition ,the overall
check of monitor, including the safety check such as the
leakage current ,should be performed only by qualified
personnel once every 12 months.
Necessary maintenance must be performed by
qualified service engineers ONLY. There are no user
serviceable parts and users are not permitted to
maintain it by themselves.
The oximeter cannot be used together with devices not
specified in User’s Manual.Only the accessory that
appointed or recommendatory by manufacture can be
used with this device.
This product is calibrated before leaving factory.
1.2 Warning
Explosive hazardDO NOT use the oximeter in
environment with inflammable gas such as some
ignitable anesthetic agents.