18
Telehealth medical diagnostic system IDIS2GO
Appendix I
Instructions and declaration of the manufacturer - electromagnetic emissions -
for all DEVICES AND THE SYSTEM
Instructions and declaration of the manufacturer - electromagnetic radiation
The SP10W is intended for use in the electromagnetic environment specified below. The user of the SP10W
must ensure that it is used in such an environment.
Electromagnetic environment
SP10W uses RF energy only for internal function.
Therefore, its RF emissions are very low and probably do
not cause any interference in the neighbouring electronic
equipment.
SP10W is suitable for use in all institutions, including
domestic institutions and those directly connected to the
public low-voltage power supply network in buildings,
which is used for indoor needs.
Instructions and declaration of the manufacturer - resistance to electromagnetic interference -
for all DEVICES AND THE SYSTEM
Instructions and declaration of the manufacturer - resistance to electromagnetic interference
The SP10W is intended for use in the electromagnetic environment specified below. The user of the SP10W
must ensure that it is used in such an environment.
Electromagnetic
environment (instruction)
Electrostatic discharge
(ESD)
IEC 61000-4-2
Floors should be wooden,
concrete or made of
ceramic tile.
If the floor is covered with
synthetic material, the relative
humidity should be at least
30%.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
The quality of the power
supply must be
standardised,
commercial or hospital
environment.
Instructions and declaration of the manufacturer - resistance to electromagnetic interference -
for all DEVICES AND THE SYSTEM that do not need POWER
Instructions and declaration of the manufacturer - resistance to electromagnetic interference
The SP10W is intended for use in the electromagnetic environment specified below. The user of the SP10W
must ensure that it is used in such an environment.
Electromagnetic environment (instruction)
RF radiation
IEC61000-4-3
The device for manual or mobile radio
communication should be used no closer to any
part of the SP10W, including cables, than from
the recommended distance, which is calculated
from the equation applied to the transmitter
frequency.
Recommended separation distance