5-3
Adaptis A1000
Incubator
230V/50Hz 120V/60Hz
Kit total inputs amp 1.42 2.7
Light average* 115 115
Light average with highlight accessory* 200 200
Number of fixtures 4 4
Cabniet total Input Amps 5.9 10.6
Total input amps 7.3 13.3
Minimum total circuit ampacity 8 14.9
Maximum overcurrent protection 15 20
Additional fixture 0.35 0.65
Adaptis A1000
Arabidopsis
230V/50Hz 120V/60Hz
Kit total inputs amp 2.83 5.4
Light average* 300 300
Light average with highlight accessory* n/a n/a
Number of fixtures 2 2
Cabniet total Input Amps 5.9 10.6
Total input amps 8.7 15.8
Minimum total circuit ampacity 9.4 17.4
Maximum overcurrent protection 15 20
Additional fixture 1.42 n/a
Adaptis A350
Ararbidopsis
230V/50Hz 120V/60Hz
Light average* 280 280
Number of fixtures 3 vertical 3 vertical
CABINET Total Input Amps 8.4 15.1
TOTAL INPUT AMPS 8.4 15.1
Minimum TOTAL circuit ampacity 9.1 16.4
Maximum OVERCURRENT PROTECTION 15 20
Additional fixture n/a n/a
*Light measurement at 6” (150mm), chamber and ambient temperature of 25°C.
Light intensities are nominal values measured at the rated chamber supply voltage.