May 2009 | Revision 6.0
1-4
CHAPTER 1.0 | PRECAUTIONS
1.2 Installation precautions
Preparation
Electrical Connections
Drain Connections
Read these instructions carefully before
proceeding. Ensure building power to the
chamber is off prior to installation.
The Adaptis A350 reach-in chambers are
shipped in a single crate while the A1000
chambers are generally shipped in two crates
one for the cabinet and another for the kit. To
avoid damage, keep the chamber crated until
it is ready for placement within the facility.
Before moving the chamber to its final
location, make sure it clears through
doorways, hallways, elevators, etc. You may
need to uncrate it before moving it. The door
may be removed if absolutely necessary.
Before connecting the chamber to building
electrical service, verify it matches the
service specified on chamber serial plate.
See Chamber Start-up section for details
Ensure a proper ground wire connection
from the building panel.
Verify neutral to ground voltage is within
tolerance. <3VAC.
If the unit does not contain a suitable plug,
install an electrical adapter ensuring the
presence of the connection to ground, or
replace the provided plug by one that meets
the local electrical requirements for your
chamber by qualified personnel.
A male 1/2” NPT (A1000 and A350) or a 1”
OD hose (A350) gravity drain is under the
rear of the chamber. Connect to either an
open drain or a tapped gravity drain.
If your equipment has been ordered with
the DP option (drip pan) there is no need to
connect the drain