1-3
Uncrating & Unpacking
Environmental Conditions
Shelving
Positioning
Replacing Kits
(only for A1000 models)
1. Using a utility knife, carefully uncrate
the chamber.
2. Verify receipt of all components per
the packing list. Notify your Dealer or
Conviron Sales member i immediately of
any missing components.
3. Ensure all units’ doors are locked
before moving the chamber to
the installation location.
4. Remove kit parts from packaging
where necessary.
Conviron guarantees performance of
the chamber for ambient temperature
conditions between 15°C and 30°C. It is
important to ensure that the room in which
the chamber is located adheres to these
environmental conditions.
Chambers with air-cooled condensing units
need a minimum of 1 foot (300mm) of space
at the back for proper operation.
Adaptis chambers come with wire shelves.
These shelves do not slide over their
supports. To position the shelves, simply
locate the supports according to the
required shelf height and gently lower the
wire shelf over the supports. Push them
completely back into position.
Once the chamber is moved into position,
ensure that the levelers (4 in total) are
unscrewed until making firm contact with the
floor. Ideally, the levelers in the front should
be adjusted slightly higher than those in the
back, to facilitate drainage.
When a kit is installed in the cabinet for the first
time, all four lamp connectors inside the cabinet
are capped. Adaptis cabinets can operate
at high temperature and humidity creating
a hazard if the connector is exposed to such
environment. Keep the caps on the unused
connectors at all times. When replacing kits
ensure to re-cap all connectors not being used.