3-3
Water supply requirements
• Flow:Maximum:0.25gallons/hour(0.9Litres/hour)
• Waterquality:Aconnectiontoawatersourceisrequiredforchambers
• Withadditivehumidity.
• pH: 7.0±0.5
• Filtration: <2micronsor0.00008inch
• Purity: Resistance 0.5 to 5.0 Mega Ohms or Conductivity 2.0 to 0.2µSiemens
Note: It is important to use a water source with the quality stated above, as failure to
do so will void the product warranty.
Installing your kit
See previous Section of Operator’s Manual.
3.2 Start-up
The unit is now ready to operate. The breaker, located inside the machine compartment,
should be in the ON position. The main power switch is located on the left hand side of the
unit between the top panel and the door. Open the door slightly and push the button once.
The switch is backlit and turns red for power ON. (See figure 3.4.)
In a few seconds the Controller will boot up.
The user now has to configure some parameters in the controller to adequate the
controller’s response to the specific model being setup.
Find how to adjust these parameters in the Operator’s Manual, within the options menu.
Figure 3.4: Main power switch.