2
1
Thank you for purchasing the COXO C-Sailor .
We recommend that prior to use, you read this document carefully regarding instructions for using, handling
method, or maintenance check so that you can carry on using the unit in the future.
In addition, please keep this operation manual in a place where a user can refer to it at any given time.
• Type of protection against electric shock :
-Class l equipment
• Degree of protection against electric shock :
-Type B applied part
• Method of sterilization or disinfection recommended by the manufacture :
-See 7. Cleaning, disinfection, packing and sterilization
• Degree of protection against ingress of water as detailed in the current edition of IEC 60529 :
-Foot Control: IPX7 (Protected against the effects of 30 minutes immersion in water)
Mode of operation :
-Intermittent operation(ON 2 min/OFF 10 min)
dental implant motor system
Classification of equipment
-The Implanted is designed for intermittent operating mode with an operating time of 2 minutes and an idle time
of 10 minutes. If the operating mode specified is observed no overheating of the system and therefore no injury
to the patient, user or third persons arises.
-Not to position the ME equipment so that it is difficult to operate the disconnection device.
1. Safety precautions prior to use
C-Sailor is intended for use in dental oral surgery and surgical procedures by qualified personnel.
INTENDED TO USE
The system may present a possibility of malfunction when used in the presence of an electromagnetic interference wave. Do not
install the system in the vicinity of the device which emits magnetic waves. Turn off the power switch of the Control Unit of this
system when an ultrasonic oscillation device or an electrode knife is located in the vicinity is used.
• C-Sailor needs special precautions regarding EMC and needs to be installed and put into service according to the EMC
information.
• Portable and mobile RF communications equipment can affect C-Sailor. Do not use RF equipment outskirts for the product.
• The use of accessories, Motors and cables other than those specified, with the exception of Motors and cables sold by the
manufacturer of C-Sailor as replacement parts for internal components, may result in increased EMISSIONS or decreased
IMMUNITY of the Control Unit.
• C-Sailor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the
Control Unit should be observed to verify normal operation in the configuration in which it will be used.
• When operating this system always consider the safety of the patient.
• Read this Operation Manual before use, and fully understand the functions of each part for operation key.
• Do not attempt to disassemble the Control Unit /Foot Control / Micromotor nor temper with the mechanism.
• Check for vibration, noise and overheating before use and if any abnormalities are found in use, stop using immediately and
contact your dealer.
• Use an electrical outlet that is grounded.
• Do not drop, hit, or subject to excessive shock.
• Do not bend the Irrigation Tube while the water pomp is operating. It could cause tube breakage.
• Do not use bent, damaged or sub-standard burs. The Shank could bend or brake.
• Do not exceed the recommended speed.
After each Operation, be sure to lubricate and sterilize the handpiece as soon as it's cleaned. Blood coagulation can cause
corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause damage.
• The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit and the Foot
Control cannot be sterilized by any method.
• When the product is very frequently used, consider the maintenance of a small stock of replaceable parts.
• Do not disconnect the Motor Cord from the motor.
• Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions.
• Follow the table below for the motion time of motor and handpiece. Continuous use over a long time may cause the overheating
of the handpiece, which in turn may result in an accident.
Applied parts for patient and/or operator is Handpiece.
WARNING
CAUTION
Read Handpiece’s Operation manual carefully before use.
• For service requirements and spare parts contact your dealer.
• The use of COXO genuine pre-sterilized, disposable Irrigation tube Kit is recommended.
• Turn off the Main Power Switch after each use.
NOTICE
Indications:
Missing a tooth, multiple teeth, the teeth of all patients.
Mandibular alveolar absorb all full mouth, wearing the traditional full mouth dentures is difficult, is not good,
wear is not strong person.
Wear traditional dentures fixed differential, non-functional mucosa can not endure.
For dentures aesthetics, function special requirements.
Remaining teeth missing teeth around the problem, can not do the traditional dentures.
Explains an instruction where minor to medium injury or physical damage may occur.
Explains an instruction where personal injury or physical damage may occur.
Explains an instruction that should be observed for safety reasons.
Classification
Degree of Danger or Damage and Seriousness
WARNING
CAUTION
NOTICE
Cautions for operation
Read these safety cautions thoroughly before use and operate the product properly.
These indicators are to allow you to use the product safely, prevent danger and harm to you and others.
These are classified by degree of danger, damage and seriousness. All indicators concern safety, be sure to
follow them.
Contraindications:
Systemic diseases (cancer, cardiovascular diseases serious diseases, the blood system, the immune system
The disease, ...).
Ongoing and topical treatment of certain systems (anticoagulant therapy, chemotherapy, radiotherapy, ...).
Poor quantity and quality of bone.