C-SMART
-I
Pro
U11er Manual
9.2
Cleaning,Disinfection
and
Sterilization
(according
to DIN EN ISO
17664)
The
procedure
for
cleaning
,
disinfection
and
Sterilization applies
only to
the
Lt.
WARNINGS
C-SMART·IPro Uoer
Manual
accessories
lip hook, file clip and
contra-angle.
9. 2.
1 Pre-treatment
Pulp and
dentin
residues
must be
removed immediately
from the
accessories
(within maximum
2 hrs). Do not let them
dry!After
the
accessories
have
been
used on
patients,
place the
accessories
for
cleaning,pre-disinfection
and
interim storage directly
into a bowl filled with an
appropriate cleaning
and
disinfecting solution
(for max. 2 hrs). Then clean the
accessories under
running
water or clean in a
disinfecting solution
to remove all
visible
contamination.
The
disinfectant should
be
aldehyde-free (aldehyde
fixes
blood stains),tested
for
effectiveness
(e.g. CE
mark),
suitable for
accessories disinfection
and
compatible
with the
accessories. Only
use
clean
,
soft
brushes
to
manually
remove
contamination
or a
clean,
soft cloth which you only use for this
purpose.
Do not use metal
brushes
or
steelwool.
For
better cleaning
of the inner parts, the file clip must be
pressed
and
released
five times during
cleaning process. Please
note
that
disinfectants
used for
pre-treatment
are only for
personal
protection
and
do
not
replace disinfection
when
cleaning
is
completed.
The
pre-treatment
process
should be
performed
in every
case.
.&.
CAUTION
Do not use an
automated
procedure
or
ultrasonic
bath to
clean
or
disinfect
the
accessories.
9.2.2 ManualCleaning
and
Disinfection
Cleaning
•
Place the
pre-cleaned
accessories
into the
cleaning
bath for the
prescribed
contact
time, the
accessories
must be
sufficiently covered
(if
necessary careful
brushing
with a soft
brush).
For better
cleaning
of the inner
parts,the
file
clip
must be
pressed
and
released
five times
during cleaning process.
·Then
remove the
instruments
from the
cleaning
bath and rinse them
thoroughly
with water for at least three times for 1 min. and press and
release
the file
clip
five
times.
Disinfection
·Place
the
cleaned
and
inspected accessories
into the
disinfection
bath
for
the
prescribed contact
time; the
accessories
must be
sufficiently covered.
For better
disinfection
of the inner
parts,
the file clip must be
pressed
and
released
five times
during disinfection process.
• Do not
disassemble
or alter the motor
handpiece.
• Do not
expose
the
unit,motor handpiece
or
battery charger
to any
liquid.
• Do not
expose
the device to
direct
or
indirect sources
of heat.
Operate and
store the device in a safe
environment.
• Do not drop the
device.
• In order to avoid
possible
risks due to
electromagnetic interference,
do
not
use any
electrical
medical
device
or
electrical
device of any other kind
in
close
proximity
to the
device.
• The
device
may
possibly malfunction if
used in the
presence
of
an
electromagnetic interference
wave. Do not
install
the
device
in the
vicinity
of any
device
that emits
magnetic waves.
• Do not use the device in the
presence
of free
oxygen
or
anesthetic
substances
or
flammable
products.
• None of the
device components
are
delivered disinfected
or
sterilized:
components
such as control
unit,motor
and motor cable need to
be
disinfected,the contra-angle
and Lip hook needs to be
sterilized
prior
to
first use and in
between
each
patient!
• The
plastic enclosure
is not
sealed
,
do not use any
liquid
or spray
directly
on the unit,
especially
on the
monitor
or near the
electrical
sockets
.
•
Follow
the file
manufacturer's instructions
for use of the
endodontic
files.
• The file
system
shown on the
display
must always match the file in use.
This
is of the utmost
importance
in order to
avoid misusing reciprocating
files
and
continuous
rotary
files.
II