9 Maintenance, Cleaning and
Disinfection
• Read these safety precautions thoroughly prior to
use
.
Thes
e
precautions
allow
you to use the product safely, preventing harm to you and
others.
• Refer to the WARNINGS chapter to verify any special care to exercise
before
starting to use the complete
device.
• Before changing the
contra
-
angle
or file, turn off the power of the unit.
Changing
with the power kept on may cause unintended rotation by
accidentaltouc
h
of
the
ON/OFF
key
.
• Always clean the shank
ofthe
file to be
installed.Allowing
dirt to enter the
chuck
could cause loss of
concentricity
and
deterioration
of chucking
force.
• Pay attention to the direction of the battery connector when
installing.
Forcible
setting In the wrong direction may cause damage and fluid leakage due to
a
short
circuit.
• Fully-charged
rechargeable
batteries generally discharge gradually over
time
even though the device is not used. It is recommended to recharge the
battery
just before
use
.
• When disposing of the control unit, follow the
instructions
of your
local
government for disposal, as they contain materials which may become
industrial
waste.
• This product does not consider a patient's age,
gender
,
weight or
nationality.
The manufacturer declines any
responsibility
in the case
of:
• Use of the device for applications other than those specified In the
instructions
for use and
maintenance.
• Modifications or repairs performed by persons not authorized by
the
manufacture
r
.
·Use
of
non-original
components or components other than those specified in
the
Standard Components chapter_
• File breakage due to
missus
• Accessories or device breakages due to
sterilizatio
n
:
none of the
device
components are
sterilizable
(except for the
contra-angle).
6 Adverse
Reactions
There are no known adverse
reactions.
II
9.1
General
The device is maintenance free and does not contain user serviceable
parts.
CJ='NOTE
Service and repair should be provided by factory trained service
personnel
only
.
• The surface of the device
,
test wire ,lip hook wire, motor cable and
lighting
device should be cleaned using tissue or soft cloth soaked with aldehyde
free
disinfecting and detergent solution
(bactericidal
and
fungicidal
)
.
&cAUTION
• Wipe The surface
ofthe
device
,
test wire ,lip hook
wire,motor
cable
and
lighting device with a clean
cloth,lightly
moistened with a
non-aggress
i
ve
disinfectant
.
• Do not apply any liquid or spray directly on the
device,especially on
the
display
.
• Do not use high-proof alcohol for
d
i
sinfection
.
• The accessories lip
hook,
file clip and contra-angle must be cleaned,
disinfected
and sterilized prior to each
use_
Thorough cleaning and disinfection are
essential
prerequisites
for effective
Sterilization.The
specific Instructions for
cleaning/
Sterilization
must be applied according to the
Instructions
in chapter 9.2
·cleaning,
Disinfection
and
Sterilization
•
.
In
addition,the
operating
instructions
of
the
devices used in your practice must be followed.As part of your
responsib
ili
t
y
for
accessories
sterility
,
always ensure that only
val
i
date
d
methods for
cleani
n
g
/
dis
i
nfect
i
on
and
Sterilization
are used, that devices
(dis
i
nfector,sterilize
r
)
are
regularly maintained and inspected and that the validated parameters
are
maintained with each
cycle.
• In
addition
,
always observe the validated legal regulations and
regula
t
ions
on
hygiene relating to your practice or the hospitaL This applies in particular to
the
guidelines regarding effective prion
inactivatio
n
.
• For your own safety, always wear protective gloves, glasses, mask
when
handling contaminated
accessories.