5
Precautions
9 Maintenance, Cleaning and
Disinfection
Read these safety precautions thoroughly prior to
use
.
Thes
e
precautions
allow
you to use the product safely, preventing harm to you and
others.
Refer to the WARNINGS chapter to verify any special care to exercise
before
starting to use the complete
device.
Before changing the
contra
-
angle
or file, turn off the power of the unit.
Changing
with the power kept on may cause unintended rotation by
accidentaltouc
h
of
the
ON/OFF
key
.
Always clean the shank
ofthe
file to be
installed.Allowing
dirt to enter the
chuck
could cause loss of
concentricity
and
deterioration
of chucking
force.
Pay attention to the direction of the battery connector when
installing.
Forcible
setting In the wrong direction may cause damage and fluid leakage due to
a
short
circuit.
Fully-charged
rechargeable
batteries generally discharge gradually over
time
even though the device is not used. It is recommended to recharge the
battery
just before
use
.
When disposing of the control unit, follow the
instructions
of your
local
government for disposal, as they contain materials which may become
industrial
waste.
This product does not consider a patient's age,
gender
,
weight or
nationality.
The manufacturer declines any
responsibility
in the case
of:
Use of the device for applications other than those specified In the
instructions
for use and
maintenance.
Modifications or repairs performed by persons not authorized by
the
manufacture
r
.
·Use
of
non-original
components or components other than those specified in
the
Standard Components chapter_
File breakage due to
missus
Accessories or device breakages due to
sterilizatio
n
:
none of the
device
components are
sterilizable
(except for the
contra-angle).
6 Adverse
Reactions
There are no known adverse
reactions.
II
9.1
General
The device is maintenance free and does not contain user serviceable
parts.
CJ='NOTE
Service and repair should be provided by factory trained service
personnel
only
.
The surface of the device
,
test wire ,lip hook wire, motor cable and
lighting
device should be cleaned using tissue or soft cloth soaked with aldehyde
free
disinfecting and detergent solution
(bactericidal
and
fungicidal
)
.
&cAUTION
Wipe The surface
ofthe
device
,
test wire ,lip hook
wire,motor
cable
and
lighting device with a clean
cloth,lightly
moistened with a
non-aggress
i
ve
disinfectant
.
Do not apply any liquid or spray directly on the
device,especially on
the
display
.
Do not use high-proof alcohol for
d
i
sinfection
.
The accessories lip
hook,
file clip and contra-angle must be cleaned,
disinfected
and sterilized prior to each
use_
Thorough cleaning and disinfection are
essential
prerequisites
for effective
Sterilization.The
specific Instructions for
cleaning/
Sterilization
must be applied according to the
Instructions
in chapter 9.2
·cleaning,
Disinfection
and
Sterilization
.
In
addition,the
operating
instructions
of
the
devices used in your practice must be followed.As part of your
responsib
ili
t
y
for
accessories
sterility
,
always ensure that only
val
i
date
d
methods for
cleani
n
g
/
dis
i
nfect
i
on
and
Sterilization
are used, that devices
(dis
i
nfector,sterilize
r
)
are
regularly maintained and inspected and that the validated parameters
are
maintained with each
cycle.
In
addition
,
always observe the validated legal regulations and
regula
t
ions
on
hygiene relating to your practice or the hospitaL This applies in particular to
the
guidelines regarding effective prion
inactivatio
n
.
For your own safety, always wear protective gloves, glasses, mask
when
handling contaminated
accessories.