CoCo-80 Quick Reference
64
Figure 65: Parameters menu
Analysis Parameters contains the variables that will need
to be tailored to the frequency range and measurement
quantity. The options include settings for:
Measurement Quantity: Acceleration, velocity,
displacement.
Fmax (Hz): The highest frequency that can
accurately be recorded.
Fmin (Hz): The lowest frequency that can accurately
be recorded. This must be set to between 0% and
95% of Fmax.
Number of Samples/Lines: Spectra are calculated
from discrete divisions of the time stream. Number
of samples specifies how many samples are taken
before a spectrum is calculated. The number of
Lines specifies how many bins the spectrum will
have.
Average Type: Linear or exponential averaging of
spectra.
Average Number: The number of calculated spectra
to be averaged.