INSTRUCTIONS FOR USE
MS-39 AS-OCT| IFU302EN00 - 03/2019
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
22/52
Immunity test
Electromagnetic environment
Electrostatic
discharge. IEC
61000-4-2
8kV contact.
2/4/8/15 kV
air.
IEC 60601-
1-2 Test
level
concrete or ceramic tile. If the
floors are covered with
synthetic material the relative
humidity should be at least
Radiated
electromagnetic
fields. IEC
61000-4-3
3V/m 80 MHz
to 2.7 GHz
IEC 60601-
1-2 Test
level
communication equipment
should be used no closer to any
part of the device, including
cables. Minimum distance 30
Electrical fast
transient/burst.
IEC 61000-4-4
supply lines.
1kV for
input/output
lines > 3 m
IEC 60601-
1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC
61000-4-5
differential
mode 0.5/1kV
common
mode
IEC 60601-
1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment.
Conducted
disturbances,
inducted by RF
fields. IEC
61000-4-6
3 V 150 kHz to
80 MHz
IEC 60601-
1-2 Test
level
communication equipment
should be used no closer to any
part of the device, including
cables. Minimum distance 30