INSTRUCTIONS FOR USE
MS-39 AS-OCT| IFU302EN00 - 03/2019
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
48/52
5.4 TROUBLESHOOTING
Inconvenient
Cause
Solution
Note
The device does
not switch ON
Power cable not
connected with
the power supply
Connect the
power cable of
the device to the
power supply
Press the ON/OFF
button on the
device.
If the device is
powered trough
the auxiliary
power supply of
the table, check
the connection of
the table to the
power line. Check
the functioning of
the table fuses.
The PC does not
start
Power cable not
connected with
the power supply
Connect the
power cable to
the power supply.
Push the button
of the power
supply on ON.
Replace the PC.
Make sure the
power outlet of
the room works
properly.
PC Operating
system does not
start
Hard Disk failure.
Spoiled operating
system.
Replace the Hard
Disk. Reinstall the
operating system.
Replace the PC
Make sure the
New PC features
are equivalent to
those required by
the device.