Immunity test
Electromagnetic environment
Voltage dips.
Short
interruptions
and voltage
variations on
power supply
input lines. IEC
61000-4-11
0% Un for 0.5
cycle. 0% Un
for 1 cycle.
70% Un for 25
cycles. 0% Un
for 5 s
IEC 60601-
1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it is
recommended that the device
is powered from an
uninterrupted power supply or
frequency
(50/60Hz)
magnetic fields.
30 A/m
IEC 60601-
1-2 Test
level
Power frequency of the
magnetic fields should be that
of a typical commercial or
hospital environment.