EasyManua.ls Logo

CSO MS39 AS-OCT - Page 25

CSO MS39 AS-OCT
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MS-39 AS-OCT| IFU302EN00 - 03/2019
INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
23/52
Immunity test
IEC 60601-1-2
test level
Conformity
level
Electromagnetic environment
Voltage dips.
Short
interruptions
and voltage
variations on
power supply
input lines. IEC
61000-4-11
0% Un for 0.5
cycle. 0% Un
for 1 cycle.
70% Un for 25
cycles. 0% Un
for 5 s
IEC 60601-
1-2 Test
level
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it is
recommended that the device
is powered from an
uninterrupted power supply or
battery.
Power
frequency
(50/60Hz)
magnetic fields.
IEC 61000-4-8
30 A/m
IEC 60601-
1-2 Test
level
Power frequency of the
magnetic fields should be that
of a typical commercial or
hospital environment.

Related product manuals