Cygnus 2+ Operating Manual
Wipe clean the step wedge then
add some fresh couplant to the
step wedge
Place the Probe on the sample
and verify the gauge can get a
thickness value.
Press the Up Select key to get to
the Calibrate Two Point item
Press the Menu key to start
Start with the Thick sample –
Maximum thickness.
While holding the probe firmly
on the thick sample, and while a
steady thickness measurement