Cygnus 2+ Operating Manual
Press the Up select key to proceed
Wipe any couplant from the probe face
Press the Up select key to proceed
The gauge measures the probe zero
If the Probe Zero fails a message will be
displayed.
• Check the probe has not been
unplugged or the cable is
damaged/faulty.
• The probe face must be clean and
in the air.
• Are you using a non-Cygnus
Setting the Velocity of Sound
The gauge uses the Velocity of Sound value to calculate the
material thickness value. It is therefore important the velocity
value is set for the material being measured.
If you perform a Calibration (single or two point) the Velocity
of Sound will be set for you during the calibration – so you
don’t need to adjust it afterwards.
You can manually set the velocity of sound value if required,
normally you would do this if;
• You can’t perform a calibration
• You want to use the same velocity setting as last time
• You want to use a velocity from a material list