Datex-Ohmeda F-LM1 User Manual

Datex-Ohmeda F-LM1
377 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Datex-Ohmeda
S/5
TM
Light Monitor, F-LM1, F-LMP1
Technical Reference Manual
All specifications are subject to change without notice.
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Outside the USA, check local laws for any restriction that may apply.
Order Code M1037863
October 2005
Datex-Ohmeda, Inc.
P.O. Box 7550
Madison, WI 53707-7550, USA
Tel. +1-608-221-1551
Fax +1-608-222-9147
Healthcare Finland Oy
Helsinki, Finland
P.O. Box 900
FIN-00031 GE, FINLAND
Tel: +358 10 39411
Fax : +358 9 146 3310
www.datex-ohmeda.com www.gehealthcare.com
© 2005 Copyright General Electric Company

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Datex-Ohmeda F-LM1 and is the answer not in the manual?

Datex-Ohmeda F-LM1 Specifications

General IconGeneral
BrandDatex-Ohmeda
ModelF-LM1
CategoryMedical Equipment
LanguageEnglish

Summary

INTRODUCTION

Installation

Covers choosing a location and important warnings for safe installation.

SAFETY

Safety precautions

Detailed safety instructions, including warnings, cautions, and handling advice.

Safety test

Procedures for performing electrical safety and leakage current tests.

FUNCTIONAL CHECK

General functional inspection

Basic checks for power, startup, and essential functions like speaker and battery.

Performance checks

Tests all major parameters (ECG, SpO2, InvBP, NIBP, CO2) using a patient simulator.

SERVICE CHECK

Service check procedure

Detailed steps for performing a comprehensive service check of the monitor.

ADJUSTMENTS AND CALIBRATIONS

NIBP calibrations

How to check and recalibrate the NIBP measurement for accuracy.

Temperature calibration

Steps to calibrate the temperature measurement channels using test plugs.

Invasive pressure calibration

Procedure for calibrating invasive pressure channels when transducers or boards are replaced.

GENERAL TROUBLESHOOTING

Troubleshooting flowchart

A visual guide to diagnose and resolve common monitor startup problems.

SERVICE MENU