EasyManua.ls Logo

DayTronic 3000PLUS - Page 49

DayTronic 3000PLUS
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3KP64 - 3.25
3000PLUS WITH 5D64 INSTRUCTION MANUAL V. SB.2
Daytronic Corporation
2211 Arbor Blvd. Dayton, OH 45439 (800) 668-4745
Tel: (937) 293-2566 Fax: (937) 293-2586 www.daytronic.com
SETUP STAGE 4(T): TWO-POINT CALIBRATION
(contd)
When the
3000PLUS
displays this . . . Do this . . .
4(T).h. Use
and
as explained in Section 3.A to adjust the displayed number until the
desired
NEGATIVE OUTPUT READING is obtained (correctly repre-
senting the present amount of negative load, expressed in measure-
ment units).
NOTE: You will not be allowed to change the displayed
number by more than 2% of its absolute value.
Then press
4(T).i. The 3000PLUS lets you apply a midscale POSITIVE LINEARITY COR-
RECTION
. Adjusting the 3000PLUS instruments POSITIVE LINEARITY
(LPN)
factora percentage of midscale output not greater than ±2%is
useful in cases where the output linearity error increases and decreas-
es smoothly (with no inflections) with increasing values of input, as is
commonly the case with conventional LVDT sensors. If you want to
apply the positive correction factor, answer YES by pressing
and proceed to Step 4(T).j.
If no positive linearity correction is required, answer NO by pressing
and proceed to Step 4(T).l.
4(T).j. Before you can enter the correction, you must apply input loading in the
positive direction to approximately
half of the transducers nominal full-
scale rating. When this has been done, press
4(T).k. Use
and
as explained in Section 3.A to adjust the displayed number until the
desired
POSITIVE MIDSCALE OUTPUT READING is obtained (correct-
ly representing the present amount of positive load, expressed in mea-
surement units).
NOTE: You will not be allowed to change the displayed
number by more than 2% of its absolute value.
ENTER
ENTER
ENTER
ST
and
CL
are lit.
3. FRONT-PANEL CONFIGURATION
AND CALIBRATION
alternating with the
LIVE SCALED
OUTPUT READING
(CHANNEL 1)
alternating with the
LIVE SCALED
OUTPUT READING
(CHANNEL 1)

Related product manuals