EasyManua.ls Logo

DayTronic 3000PLUS - Page 53

DayTronic 3000PLUS
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3KP64 - 3.29
3000PLUS WITH 5D64 INSTRUCTION MANUAL V. SB.2
Daytronic Corporation
2211 Arbor Blvd. Dayton, OH 45439 (800) 668-4745
Tel: (937) 293-2566 Fax: (937) 293-2586 www.daytronic.com
SETUP STAGE 5: LIMITS
(contd)
When the
3000PLUS
displays this . . . Do this . . .
5.i. Press
5.j. The instruments currently effective
LOW LIMIT setpoint value will be dis-
played (having the decimal-point precision specified in Step 2.f). The LOW
LIMIT setting and HIGH LIMIT setting (see below) define the 3000PLUSs
three distinct LIMIT ZONES, as diagramed in Fig. 21 (Section 5.E)*:
LESS THAN ZONE (LO): the reading for Channel 2 is less than the
current Low Limit
BETWEEN ZONE (OK): the reading for Channel 2 is greater than or
equal to the current Low Limit and less than or equal to the current
High Limit
GREATER THAN ZONE (HI): the reading for Channel 2 is greater
than the current High Limit
The LESS THAN or GREATER THAN zones may be effectively extended
for
nonlatching limits by means of a user-specified HYSTERESIS DEAD-
BAND (see below).
Use
and
as explained in Section 3.A to adjust the displayed number until the
desired limit value is obtained (expressed in measurement units).
NOTE:
You will not be allowed to enter a LOW LIMIT that is greater than the cur-
rent HIGH LIMIT (see below).**
Then press
5.k. Press
5.l. The instruments currently effective
LOW LIMIT HYSTERESIS value will be
displayed (having the decimal-point precision specified in Step 2.f). This
parameter lets you define a HYSTERESIS window (or deadband) immedi-
ately above the LESS THAN LIMIT ZONE defined by the current LOW
LIMIT setpoint value (see Fig. 21, Section 5.E). The hysteresis deadband is
to prevent low-level signal noise from toggling the low-limit relays on and
off while the reading of Channel 2 remains in the neighborhood of the set-
point.
(contd)
ENTER
ENTER
ENTER
ST
and
LM
are lit.
3. FRONT-PANEL CONFIGURATION
AND CALIBRATION
* For a complete discussion of limit
monitoring, see Section 5.E. For
front-panel limit status indication,
see Section 1.D.
** Also, the absolute value of the
LOW LIMIT value should not be
greater than that of the
FULL-
SCALE OUTPUT IN ENGINEER-
ING UNITS
entered in Step 2.i.

Related product manuals