EasyManua.ls Logo

DayTronic 3000PLUS - Page 56

DayTronic 3000PLUS
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3KP64 - 3.32
3000PLUS WITH 5D64 INSTRUCTION MANUAL V. SB.2
SETUP STAGE 6: AUXILIARY OUTPUT
When the
3000PLUS
displays this . . . Do this . . .
6.a. Press
6.b. The meters current
PEAK CAPTURE MODE will be displayed: POSITIVE
PEAK (P PEAK) or NEGATIVE PEAK (N PEAK). Here, positive peak
operation refers to the analog capture and hold of the
maximum (most
positive) excursion of the auxiliary DAC output (Channel 2), while nega-
tive peak operation refers to the capture of Channel 2s
minimum (most
negative) excursion. The 3000PLUS will be placed in the selected mode
whenever peak capture is enabled by means of a logic input at the rear
PEAK terminal (for a complete discussion of peak capture operation, with
diagrams, see Section 5.B; for LOGIC I/O connections, see Section 2.E).
Press
to toggle between the two allowed modes. When the desired peak cap-
ture mode is displayed, press
6.c. Press
6.d. The instruments current
PEAK DEFEAT THRESHOLD will be displayed
(having the decimal-point precision specified in Step 2.f). By means of this
parameter, you can set up a peak-defeat input threshold in order to pre-
vent induced low-level signal noise from triggering a HAVE PEAK condi-
tion, when analog peak capture is enabled (see Fig. 18, Section 5.B).
Within the low-level deadband defined by this threshold, the output will
simply track the input, regardless of signal behavior.
Use
and
as explained in Section 3.A to adjust the displayed number until the
desired peak defeat threshold value is obtained (expressed in measure-
ment units).
NOTE: You will not be allowed to enter a negative number.*
Then press
(contd)
ENTER
ENTER
ENTER
SCROLL
ENTER
ST
and
PK
are lit.
3. FRONT-PANEL CONFIGURATION
AND CALIBRATION
PEAK MODE
SELECTIONS:
P PEAK
N PEAK
Daytronic Corporation
2211 Arbor Blvd. Dayton, OH 45439 (800) 668-4745
Tel: (937) 293-2566 Fax: (937) 293-2586 www.daytronic.com
* Also, the threshold value should
not be greater than 20% of the
FULL-SCALE OUTPUT IN ENGI-
NEERING UNITS
entered in Step
2.i.

Related product manuals