Copyright © 2005 Data Design Corporation Version 05.02.06
All rights reserved.
31
4.0 Calibration
The TR122 calibration compensates for variability in the analog instrument electronics.
The instrument is delivered fully calibrated and will not need any calibration efforts during
normal use. Calibrations are generally good for at least one year.
There are two aspects to the calibration of the TR122. The first is a familiar process
where the analog front end frequency compensation is adjusted to provide the best possible
representation of the incoming signal. This calibration step involves actual adjustment of
compensation capacitors in the front end electronics, similar to adjusting the compensation on
an oscilloscope probe. The other aspect involves compensation for gains and offsets of the
instrument electronics. This information will be discovered during calibration and stored in
a non-volatile memory in the TR122 as opposed to making any physical adjustments.
The primary objective of calibration is to be sure that transient record data displayed
or stored to disk matches the signal being measured within the limits of the instrument. For a
transient recorder, once the front end frequency compensation is adjusted the only other items
to compensate for are any irregularities in gain or offset present in the analog instrument
electronics. The gain calibration information stored in the instrument will be used to adjust
data retrieved from the instrument before it is presented on the screen or saved to disk. This is
an automatic function of the API library, so it will be present in any application which uses
this library.
In addition, the non-volatile memory is used to store information about the effect
of offset voltages on the instrument electronics and what offset is present initially. This
information is used to adjust the offset voltage applied such that zero offset is really zero offset
and any requested offset matches the intent. This information is applied every time the offset
is adjusted. The offset calibration tends to be different between settings of the instrument
attenuator electronics. Therefore, the API automatically reapplies the offset when the attenuator
settings are changed.