User_Manual-PILATUS-V1_2.doc 19/57
Command or Macro Description
menu Shows all commands
It is divided in 4 parts:
TVX reserved words
TVX defined commands
Macros, saved in default.gl or typed by the user
Variables
help command -or-
man command
Shows the help text for the command. Help help is a
good way to start. Help “gr*” will show all commands
beginning with ‘gr’; the quotes are required, else the
system takes ‘*’ to mean multiplication.
ESC-button Stop a running task and return to the TVX line
interpreter
CTRL-C Full reset of TVX
cam k stops the running Camserver processes
rbd Read Back Detector.
Self test of the digital part of the detector. Sends a
digital pattern to each pixel, reads it out and displays
the image.
Use this command always after a startup.
Every pixel should show 1000 counts.
calibdet Self test of the detector.
Sends 100 calibrate pulses to the analog part of the
detector, reads back the recorded values as an image
and displays the result. Every pixel should show 100
counts.
Use this command always after a startup.
Should the pixels not show 100 counts in the absence
of radiation, repeat the command. If the image is
black, type setdac, then repeat the command. In case
this tests fails, turn the power supply off, close the
TVX and Camserver windows and start up again.
setdac Sets all Digital Analog Converters (DAC) to the
predefined values.
imagepath path
Image Path
Without the input of a path it displays the current
default path. With a declaration it changes the default