User_Manual-PILATUS-V1_2.doc 40/57
Figure 19 Distribution of the trimbit settings
Trimming of the module at the current X-ray energy is now complete. You are
asked if you would like to load the trim settings you have just taken for this
module. Set the variable flag to equal 1 to do this and the load trim glossary
will be started: flag=1.
For standard settings, a faster way to generate trim-files is to use
the following table for the recommended values for Vtrim and Vcmp
Settings for Vrf = -0.2
E [keV]
Vcmp0 [V]
Vtrim [V]
6 0.75 1.38
8 0.6 1.37
10 0.44 1.36
12 0.3 1.35
Table 1: Recommended Values for Vcmp and Vtrim for standard settings and
energies between 6 and 12 keV.
These values can be used together with the glossary trimscan_only.gl, where
you directly enter Vtrim and Vcmp for the targeted energy.