Loading...
10 | Safety Information
EMC notice
Electromagnetic emissions
The MR2217 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MR2217 should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment - Guidance
RF emissions
CISPR 11
Group 1 The MR2217 uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are
not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B The MR2217 is suitable for use in all
establishments, including domest
ic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicke
r emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
Immunity test IEC 60601 Test
levels
Compliance
level
Electromagnetic environment
- Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
tr
ansie
nt/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
outp
ut lines
±2 kV for power
supply lines
±1 kV for input/
o
utput lines
Mains power quality should be that
of a
typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(
s
)
±2 kV line(s) to
eart
h
±1 kV line
(s) to
lin
e(s)
±2 kV line(s) to
ea
rt
h
Mains power quality should be that
of a typical
commercial or hospital
environment.

Table of Contents