8555-M512-O1 Rev E y MedVue MV1 Owner’s 30
CALIBRATION SETUP, CONT.
Multi-point Calibration
C
CAL WEIGHT 1
(First Calibration Weight)
The display will show CAL WEIGHT 1 (which is a prompt for entry of the first calibration
weight value) and placement of this amount of test weight on scale platform.
NOTE: The MV1 defaults to half (½) of the scale capacity.
1. Make certain scale platform is empty and free of debris.
2. When the display is showing CAL WEIGHT 1, the current setting for the first calibration
weight will be displayed also. If the value displayed is acceptable, place the calibrated
test weight on the scale platform, and then press the ENTER key. Proceed to Step 5.
3. Otherwise, determine the exact amount of test weight to be placed on the scale platform,
and then use the numeric keys to enter this value. Allowable values are 1 through scale
capacity (refer to the capacity label on MV1).
4. Verify that the numbers entered are the same as the amount of test weight, and then
press the ENTER key.
5. The display will change to show PLEASE WAIT and starting at the left and proceeding
right, a series of dashes will appear on the display, and then starting at the right and
proceeding left, disappear. Next, the display will change to CAL WEIGHT 2.
CAL WEIGHT 2
(Second Calibration Weight)
The display will show CAL WEIGHT 2 (which is a prompt for entry of the second calibration
weight value) and placement of this amount of test weight on scale platform.
NOTE: The MV1 defaults to full scale capacity.
1. When the display is showing CAL WEIGHT 2, the current setting for the second
calibration weight will be displayed also. If the value displayed is acceptable, place the
calibrated test weight on the scale platform, and then press the ENTER key. Proceed to
Step 4.
2. Otherwise, determine the exact amount of test weight to be placed on the scale platform,
and then use the numeric keys to enter this value. Allowable values are 1 through scale
capacity (refer to the capacity label on MV1).
3. Verify that the numbers entered are the same as the amount of test weight, and then
press the ENTER key.
4. The display will change to show PLEASE WAIT and starting at the left and proceeding
right, a series of dashes will appear on the display, and then starting at the right and
proceeding left, disappear. Next, the display will change to UNLOAd SCALE.
NOTE: If the same weight is used for
CAL WEIGHT 1
and
CAL WEIGHT 2
, then
the scale reverts to a Single Point calibration.
UNLOAd SCALE
(Unload Scale)
With the display showing UNLOAd SCALE:
1. Remove the weight from the scale platform.
2. Press the ENTER key.
3. The display will change to show PLEASE WAIT and starting at the left and proceeding
right, a series of dashes will appear on the display, and then starting at the right and
proceeding left, disappear.
4. Next, the display will change to show the GRAVITY ENAbLE prompt.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Detecto MedVue MV1 and is the answer not in the manual?

Detecto MedVue MV1 Specifications

General IconGeneral
BrandDetecto
ModelMedVue MV1
CategoryAccessories
LanguageEnglish