8555-M512-O1 Rev E y MedVue MV1 Owner’s 31
CALIBRATION SETUP, CONT.
G
GRAVITY
With GRAVITY displayed, press the ENTER key. The display will change to GRAVITY ENAbLE.
Proceed to the GRAVITY ENAbLE (Enable Gravity Constant) parameter.
Otherwise, press the ID/HEIGHT key to advance to the
LOCK menu.
GRAVITY ENAbLE
(Enable Gravity Constant)
The Detecto MedVue
®
Model MV1 Weight Analyzer is equipped with a gravity constant
function which means that it can be calibrated in one location and then adjusted to match
the gravity at the location where it will used.
The display will show GRAVITY ENAbLE and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise, use the numeric keys 0/NO or
1/YES to select a new setting, and then press the ENTER key to save it.
Allowable values are YES or no.
YES = Enable Gravity Constant and
proceed to CAL GRAVITY parameter
no = Disable Gravity Constant and
proceed to the HOLd MOdE parameter
CAL GRAVITY
(Calibrated Gravity ConstantOnly if GRAVITY ENAbLE = YES)
This is the gravity constant of the location where the scale was calibrated.
NOTE: Consult the factory Tech Support for the Gravity Constant for your location.
The display will show CAL GRAVITY and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a
new value, and then press the ENTER key to save it.
Allowable values are 0 through 99,999.
OP GRAVITY
(Operating Gravity ConstantOnly if GRAVITY ENAbLE = YES)
This is the gravity constant for the location where the scale will be operated.
NOTE: Consult the factory Tech Support for the Gravity Constant for your location.
The display will show OP GRAVITY and the current setting. If the value displayed is
acceptable, press the ENTER key to save it. Otherwise use the numeric keys to enter a
new value, and then press the ENTER key to save it.
Allowable values are 0 through 99,999.