4. Maintenance
A. Initial Inspection—Equipment Provider
DeVilbiss recommends equipment inspection upon delivery
Power up the DV54/DV53 PAP using AC power.
Test pressure accuracy using an outlet cap (DV54/
DV53D-620), a calibrated pressure gauge and the proce-
dures listed under Pressure Accuracy Test.
Test the keys on the keypad using the procedures listed
under Keypad Test.
B. Routine Maintenance—Patient
Wipe case with damp cloth every few days
Check air-inlet fi lter every 10 days and clean as needed
Clean air supply tubing daily
Clean mask and headgear per manufacturer’s instructions.
C. Required three years Maintenance
Equipment Provider
DeVilbiss requires the following maintenance from the equip-
ment provider:
Test pressure accuracy using procedures listed under Pres-
sure Accuracy Test. If the device is out of calibration, follow
corrective procedures as listed.
Instruct patient on fi lter maintenance described in Instruc-
tion Guide:
Standard Filter
o Inspect fi lter every ten days
o Clean as needed
o Change every 6 months
Optional Fine Particle Filter
o Inspect fi lter every ten days
o Replace if dirty
o Change every 30 days