PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 26
whenever PPC3 is in its MAIN RUN screen. Table 5 summarizes the operation of the direct
function keys and Sections 3.3.1 through 3.3.10 detail their operation.
Table 5 provides a brief summary of direct function key operation. It may be useful to
keep a copy of this summary near the PPC3, especially when first becoming acquainted with its
operation.
Table 5. Summary of PPC3 Function Key Operation
DIRECT FUNCTION KEYS ARE ACTIVE FROM THE MAIN RUN SCREEN
See corresponding manual Sections 3.3.1 through 3.3.10 for full detail.
Adjust DUT fluid head calculation
height. Set to zero to defeat
correction.
View active range and toggle
through available ranges. [ENT] on
a range activates it.
Run automated DUT purge routine
using SPLT accessory.
Select pressure unit of measure for
the active range. Choice of units
available in this menu can be
customized.
Run an automated leak check at
the current pressure.
Select the pressure measurement
mode for the active range (absolute,
gauge or negative gauge).
Set up and optimize all aspects of
PPC3 measurement, control and
displays for a specific operating
range and measurement mode.
Menu of less commonly used internal
functions and settings.
Search for and initialize internal
and external Q-RPTs available to
the PPC3 system.
Menu of commonly used setup
features including save/delete
AutoRange ranges, display
resolution setting, jog step
adjustment, upper/lower limit setting,
AutoTest data viewing, pressure
control mode selection and limits
adjustment, external 12V driver
actuation
Set up and run automated test
sequences.
Input a pressure control set point
command from the run screen.
ENTER values when editing.
3.3 DIRECT FUNCTION KEYS
3.3.1 [RANGE]
PURPOSE
To view and/or change the active pressure measurement range and associated settings.
[RANGE] selects from existing ranges. See [AutoRange] to create ranges (see Section
3.3.4).
PRINCIPLE
A PPC3 range is defined as a pressure span and associated settings (see Section 3.2.5).
A PPC3 may have multiple ranges. Each Q-RPT (and the utility sensor if present) available
to the PPC3 system has a default range which is its maximum full scale. Additional ranges,
lower than the Q-RPT’s full scale, may also be created using AutoRange (see Section 3.3.4).

Table of Contents