3. OPERATION
Page 67 © 2002-2007 DH Instruments, a Fluke Company
Table 9. Reset – Sets
RESET RESULT SEE SECTION
[UNIT] Returns pressure unit of measure to first of six available. 3.3.2
[MODE] Returns measurement mode to the natural Q-RPT mode. 3.3.3
[RANGE] Hi Q-RPT or utility sensor default range. 3.3.1
[HEAD] 0 cm height and Nitrogen medium. 3.3.7
Control Mode Dynamic 3.4.6
Control Limits Default values of Hi RPT default range. 3.4.6.1
Upper Limit Default value of Hi RPT default range. 3.4.4
Resolution 0.001% FS of Hi RPT default range. 3.4.2
Purge Disable purge function. 3.5.7.4
AutoZ AutoZ ON. P
offset
value not affected. 3.5.1
Control Ref Auto. 3.5.7.2
Leak Check 15 second run time 3.3.9
Screen Saver 10 minutes to activation. 3.5.5.1
Key Sounds Medium 3.5.5.2
Lo Vnt Automatic. 3.5.7.5
Config Return to default control coefficients. 3.5.7.1
3.5.4.2 <2UNITS
PURPOSE
Sets the six pressure units available under the UNIT function to the SI or US
default selections depending on whether the PPC3 has been factory set for SI or US
(see Section 3.3.2).
Sets the user defined unit to 1.000/Pa (see Section
3.5.6).
Sets the reference temperature for inWa unit to 20°C.
3.5.4.3 <3ATEST>
PURPOSE
• Resets Quick Test and Q-DUT Tests to default characteristics (see Section 3.3.6).
• Clears AutoTest data log (see Section 3.4.5).
3.5.4.4 <4CAL>
PURPOSE
The Reset - Cal function will reset Q-RPT, utility sensor and barometer
calibration coefficients and settings and reset AutoZ values to zero.
This will change the PPC3 calibration and could cause it to make out of
tolerance measurements.
Clears all user values affecting the calibration of Q-RPTs, utility sensor and the
on-board barometer. Does not clear factory coefficients. The Reset – Cal resets
are itemized in Table 10.