ADCS-601™
Air Data Calibration Standard
Operation and Maintenance Manual

Table of Contents

Questions and Answers

  • R
    Ronald MckaySep 8, 2025
    Why PG7601 motorized rotation will not engage in DH Instruments ADCS-601?
    • B
      beannancySep 8, 2025
      If the PG7601 motorized rotation will not engage on your DH Instruments Test Equipment, the motorized rotation system may be recovering from an overload, or the measurement made is absolute by vacuum and the reference vacuum is not under the Ready limit. Wait 25 seconds for the system to recover.
  • N
    Nicole RamosAug 23, 2025
    What to do if ADCS component displays or in DH Instruments ADCS-601?
    • D
      Dawn HahnAug 23, 2025
      If the ADCS component of your DH Instruments Test Equipment displays or where a numerical value should be, it means the number to be displayed is too large for the allocated space. This is usually due to an erroneous setting or measurement causing an out of limit high value to be calculated. Check the settings that may be causing an out of limit high measurement and adjust if necessary.
  • K
    Kristin WebbAug 25, 2025
    Why is automated pressure generation extremely slow in DH Instruments ADCS-601 Test Equipment?
    • J
      john53Aug 26, 2025
      If automated pressure generation in your DH Instruments Test Equipment is extremely slow or the pistons never float, the test volume might be too large. Try reducing the test volume.
  • G
    Gail WongAug 27, 2025
    What does it mean when PG7601 displays or in DH Instruments ADCS-601 Test Equipment?
    • T
      Todd BrownAug 27, 2025
      If the PG7601 displays or in DH Instruments Test Equipment, this indicates a communications problem with an external device or pressure generation/control component. Check the setup and communications with the external vacuum gauge or pressure generation/control component.
  • B
    Brandon HollowayAug 27, 2025
    How to fix vacuum pumps that will not run on DH Instruments ADCS-601 Test Equipment?
    • G
      George MontgomeryAug 28, 2025
      If the system’s vacuum pump(s) of your DH Instruments Test Equipment will not run, the fuse on the AC2 or AC3 power supply connection may have blown. Replace the fuse on the AC2 or AC3 IEC power connector on the rear of the Control Cabinet. The fuse should be a 250 VT 10 A.
  • A
    Anthony CalderonSep 22, 2025
    What does mean in DH Instruments Test Equipment?
    • D
      dean05Sep 22, 2025
      If ADCS Tools displays for your DH Instruments Test Equipment, the FPG8601 is in a condition in which it cannot be used. View the FPG8601 Run Screen in ADCS Tools and check for noisy output or other abnormal conditions.
  • A
    Aaron Green IISep 16, 2025
    Why is FPG8601 pressure indication unstable in DH Instruments ADCS-601?
    • A
      Angela BarnettSep 16, 2025
      If the FPG8601 pressure indication is unstable when the FPG8601 is bypassed (vented) in your DH Instruments Test Equipment, the piston-cylinder may need to be cleaned. Clean the piston cylinder.
  • J
    Jonathan WarrenSep 18, 2025
    What to do if PG7601 piston position readings seem incorrect in DH Instruments ADCS-601 Test Equipment?
    • Z
      zcoxSep 18, 2025
      If the PG7601 piston position readings seem incorrect in your DH Instruments Test Equipment, the piston position reading system may need to be adjusted. Execute the piston position detection adjustment procedure.
  • M
    meganperezSep 7, 2025
    How to brighten a dim display on DH Instruments Test Equipment?
    • M
      mossdavidSep 7, 2025
      If the PPC3, RPM4 or PG7601 front panel display is dim on your DH Instruments Test Equipment, the screen saver option may have been activated. Press any key to resume full screen power, and adjust the screen saver activation time if desired.
  • J
    Julie ScottSep 4, 2025
    Why PG7601 automated pressure control is not starting in DH Instruments Test Equipment?
    • R
      Robert HenrySep 4, 2025
      If the PG7601 automated pressure control (GEN) is not starting after setting a target and loading mass, and the measurement mode is absolute by vacuum in your DH Instruments Test Equipment, it is because in absolute by vacuum measurement mode, before automated pressure generation begins, the reference vacuum must be below 20 Pa. Wait for the vacuum to reach less than 20 Pa, and check and correct the vacuum supply if it is inadequate.

Summary

Introduction

Specifications

Details technical specifications of the ADCS-601, covering power, pressure, and environmental requirements.

PG7601 Piston Gauge Specifications

Specifies ambient conditions, measurements, and module details for the PG7601 piston gauge.

FPG8601 General Specifications

Outlines general specifications for the FPG8601, covering ambient conditions and pressure parameters.

Pressure Measurement and Control Specifications

Defines the pressure measurement and control capabilities and parameters of the ADCS-601 system.

System Overview

Pressure Standards (PG7601 & FPG8601)

Details the PG7601 and FPG8601 pressure standards, including their modules and associated handlers.

Setup

PG7601 and FPG8601 Setup Procedures

Step-by-step instructions for setting up the PG7601 and FPG8601 systems.

Power Up, Initialization, and Verification

System Initialization and Operation Verification

Procedures for initializing and verifying the operation of the PG7601, FPG8601, and the overall ADCS-601 system.

Enter Local Gravity Value

Shut Down and Restart Procedures

Procedures for safely shutting down and restarting the ADCS-601 system.

Operation

General Maintenance and Adjustments

Maintenance of Traceability and Recalibration

FPG8601 Load Cell and Linearity Maintenance

Covers zeroing, spanning, and validating the FPG8601 load cell and its linearity.

Troubleshooting

Common Symptoms and Solutions

Provides a checklist of common symptoms, probable causes, and solutions for ADCS-601 issues.