ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 52
PG7601 uses many variables in calculating defined pressures. The sources of
the variables are determined by the current SETUP file. SETUP files are viewed,
created, edited and selected using the SETUP function accessed by pressing
[SETUP] on the PG7000 Terminal. ADCS-601 systems use SETUP file #2.
VERIFY/SELECT AUTOGEN ON
Press [AutoGen] on the PG7000 Terminal. AutoGen should be ON.
VERIFY/SELECT AUTOROTATE ON
Press [Rotate] on the PG7000 Terminal. AutoRotate should be ON.
VERIFY AMBIENT CONDITION MEASUREMENTS
PG7000 automatically measures ambient conditions and uses these conditions
in its pressure calculations.
To verify that the ambient condition measurements are operating properly proceed
as follows:
Display current ambient condition readings: Press [AMBIENT] on the
PG7000 Terminal. The ambient conditions run screen is displayed.
Verify proper ambient condition readings: Compare the ambient condition
values displayed to the actual values of ambient conditions. Refer to the
PG7601 ambient condition measurement specifications when evaluating the
ambient readings (see Section
1.1.2.1).
3.4.4 FPG8601
3.4.4.1 POWER UP
Turn the FPG8601 power ON by pressing the power ON/OFF switch on the rear
panel of the FPG8000 Terminal. Observe the display as the terminal connects
with the FPG8000 Platform, tests, initializes and goes to the main run screen.
If <….Searching…..> displays for more than 5 seconds, the communications
between the FPG8601 and the FPG Terminal are failing. Check that the
FPG8601 to FPG Terminal cable is properly installed.
If the FPG Terminal fails to reach the main run screen, service may be required.
Record the sequence of operations and displays observed and contact a DHI
Authorized Service Provider (see
Table 26).
3.4.5 START UP SYSTEM CONTROLLER AND SOFTWARE
To start up the system controller proceed as follows:
Turn ON the system controller using its power switch and wait for it to complete booting
to the normal desktop.
Run the FPG Tools program using the icon on the desktop. Once the FPG Tool
program has completed booting, select [Run], [Run Monitor]. During initialization select
<Gauge> measurement mode. Wait for initialization to complete.
Run the ADCS Tools program using the icon on the desktop. Select the [ON] button on
the main ADCS Toolbar (see Section 4.3.3). ADCS Tools initializes the complete
ADCS-601 system. Warnings that the system calibration has changed may appear.
This
is normal as ADCS Tools is writing to its database for the first time. Select [OK] in
response to these warnings.
Set system local gravity and test gas species in ADCS Tools following the instructions in
Sections 3.5 and 3.6.