ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 90
The Data Grid Run Screen is available any time a test is running for which there
is a Data File. The screen vizualizes the data taken during the test, point by
point in columns.
The Data Grid Run Screen can be opened by pressing the [Data Grid] button
on the ADCS Tools Main Tool Bar (see Section
4.3.3, Table 14).
The contents of the Data Grid Run Screen
can be customized using the [Data
Grid] tab of [Tools], [Test Options] (see Section
4.4.1.4).
Figure 43. Data Grid Run Screen
4.4.3.3 DUT/REFERENCE COMPARISON RUN SCREEN
ADCS Tools uses the engine of COMPASS for Pressure® calibration software to
manage automated testing using ADCS-601. The DUT/Reference Comparison
Run Screen supported by ADCS Tools is a COMPASS for Pressure feature. See
the COMPASS for Pressure User’s Manual for additional information on this
screen and its options. The COMPASS for Pressure manual is an on-line
manual available by pressing the [?] button, when available, in COMPASS
features. It can also be accessed on the ADCS-601 system controller in the
COMPASS for Pressure program group reached using the Windows START
button , Programs.
The Reference/DUT Comparison Run Screen is available any time a test is
running. The screen vizualizes real time DUT and ADCS-601 readings and the
error of the DUT(s) compared to the ADCS-601. The information on the display
is organized into rows and columns. The columns separate different pieces of
information. The rows separate the different devices. The first row is a label that
describes the output. The second row is always the ADCS-601 output
information. All subsequent rows are for the number of DUTs in use..
The Reference/DUT Comparison Run Screen can be opened by pressing the
[Reference/DUT Comparison] button on the ADCS Tools Main Tool Bar (see
Section
4.3.3, Table 14).
Figure 44. DUT/Reference Comparison Run Screen