ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 128
Figure 56. 10 kPa/kg piston-cylinder module lubrication chart
6.6.3.4 CLEANING THE PISTON-CYLINDER
To clean the piston-cylinder, the piston-cylinder module must first be
disassembled (see Section 6.6.3.2).
DO NOT undertake piston-cylinder module disassembly without familiarizing
yourself with Section 6.6.3.2 of this manual. Incorrect disassembly may
damage or destroy the piston-cylinder element.
There are two recommended piston-cylinder cleaning methods. Use the one
with which you have most success.
Water/Detergent Method
Of the two cleaning methods, the water/detergent method is more time
consuming but it is also more thorough.
Disassemble the piston-cylinder module following the instructions given in
Section 6.6.3.2.