2. SYSTEM OVERVIEW
Page 17 © 2005 DH Instruments, Inc.
2.2.1.1 PG7601 PLATFORM
The PG7601 Platform is the instrument base into which the piston-cylinder onto
which the masses are loaded is mounted. The platform also includes the
PG7601’s electronic measurement system.
The PG7601 Platform rear panel provides the connection to the PG Terminal,
remote communication connections and pressure connection ports.
1. COM2 (RS232) - External
vacuum gauge (Reference
vacuum CDG)
2. COM3 (RS232) – PPC3 pressure
controller, AMH automated mass
handler
3. COM1 (RS232) – System
controller (PC) communications
4. Ambient temperature sensor
5. IEEE-488 - host communications
(not used)
6. Ambient relative humidity sensor
7. Pressure ports: TEST port
(DH200), VACUUM vent port
(DH200, not used)
8. PG7000 Terminal port
Figure 12. PG7601 Platform rear panel
2.2.1.2 PG7601 TERMINAL
The PG7000 Terminal contains the PG7601 power supplies. It also provides a
local user interface through a 2 x 20 vacuum fluorescent display and 4 x 4
membrane keypad. Local commands to PG7601 are not used in normal
ADCS-601 operation. See the PG7000 Operation and Maintenance manual for
details on local operation of PG7000.
1. Fluorescent display
2. Keypad
Figure 13. PG7000 Terminal front panel