4. OPERATION
Page 97 © 2005 DH Instruments, Inc.
When the test is complete the Test Complete form displays. If automatic report
generation is enabled (see Section 4.4.2.2), the Report Editor creates a report.
4.4.6.2 RUNNING A STANDARDIZATION
When the ADCS-601 is connected to the model 3682 ADC, do not execute any
function on the 3682 that will cause it to vent pressure unless the ADCS-601 is
vented first. Causing pressure on the ADCS-601 TEST(+) port to vent when the
ADCS-601 is not vented may cause damage to the ADCS-601.
NOTE: The ADCS-601 is always vented before the model 3682 vents if the
procedures below are followed as instructed.
The model 3682 standardization procedure steps the operator through the model
3682 required sequence. This procedure cannot be fully automated due to
limitations in the model 3682 remote validation procedure.