ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 96
When prompted by the popup below, abort model 3682 operation by
pressing its red [ABORT] button. Then set <Calibration Automatic> mode
and the IEEE-488 address on the model 3682 screen as follows:
Press [C
alibration Automatic]
Press [A
dvance]
Press [I
EEE Address]
Press [2] to set 2 as the IEEE-488 address.
Press [E
nter].
The model 3682 will remain in the IEEE address selection screen until ADCS
Tools initializes model 3682 communications.
After all steps on the 3682 are complete and its display is the IEEE address
selection screen, press [OK] on the ADCS Tools popup. It is imperative that
the model 3682 be in the correct mode of operation when the test begins.
ADCS Tools sets up the the model 3682 remote interface for automatic
calibration. Ten to twenty seconds after [OK] is pressed, the model 3682
prints out the current mode of operation. ADCS Tools proceeds to run the
test automatically. The output of the model 3682 and the test progress
display on ADCS Tools as the test excutes through the test pressure points.
Use the display buttons on the ADCS Tools Main Toolbox to display other
system windows if desired.