ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 40
3.3.3 POSITION THE PG7601 AND FPG8601 PLATFORMS ON THE
REFERENCE BENCH
Place the PG7601 and FPG8601 Platforms and their Terminals in their approximate positions
on the bench (see Figure 7).
3.3.4 PREPARE FPG8601
Attach the FPG manifold to the
KF16 FPG8601 high control
port. Do not make the low
control port connection.
Connect the 1/8 in. plastic tube
included with the manifold
between valve 10 on the
manifold and the V10 port on
the FPG platform.
Attach FPG valve 11 to the 1/4
Swage fitting on the top of the
FPG manifold. Orient the
valve so that the flow arrow
points away from the manifold
(up). Connect the 1/8 in.
plastic tube included with valve
11 between valve 11 and the
V11 port on the FPG platform.
Install a KF16 blank-off cap on
the FPG8601 low test port.
Install 1/8 in. NPT M x 1/8 in. Legris adaptors into the DRIVE IN and LUBRICATION
SUPPLY ports on the rear of the FPG8601 Platform (see Figure 20).
Install a 1/8 NPT M x 1/4 in. Swage adaptor into the VACUUM port on the rear of the
FPG8601 Platform (see Figure 20).
Connect the FPG vacuum tube assembly (1/4 in. PFA with 1/4 in. Swage x KF16
adaptor installed) between the FPG VACUUM port and the KF16 tee previously
installed on the turbo pump under the bench.
Connect the FPG8000 Terminal to the rear of the FPG8601 platform rear with the 25 pin
cable.
Connect the FPG8601 power cord (16 AWG) from the FPG8000 Terminal to the power
strip underneath the bench. Do not power up the FPG8601.
Fill the FPG8601 bubbler bowl with distilled water (see Section 6.2).