3. INSTALLATION
Page 49 © 2005 DH Instruments, Inc.
3.3.11 COMPLETE REFERENCE VACUUM INTERCONNECTIONS
Connect the KF40 cross to the AMH-
38 right side (viewed from front) KF40
fitting.
Connect the KF40 cross to the
vertical vacuum tube (that comes up
through the bench). Adjust the
position of the vertical tube if
necessary by loosening and
retightening the holding collar on the
vertical tube brackets (under the
bench).
Connect the KF40 cross to the
bellows adaptor KF40 connection.
Connect the KF25 port on the bellows
adaptor to the FPG8601 valve 9 KF25
port. It may be necessary to move the
FPG slightly.
Connect the KF40 blank-off cap to the top port of the KF40 cross.
Verify the bubble level on top of the FPG8601 platform and re-level if necessary.
3.3.12 INSTALL THE UTILITY VACUUM PUMP
Position the utility vacuum pump under the bench. The two scroll vacuum pumps
provided are identical, so either one may be used here.
Install a 3/8 in. Swage x 1/4 in. NPT M adaptor on the Control Cabinet UTILITY VACUUM port.
Connect the utility pump inlet to the Control Cabinet UTILITY VACUUM port using the
3/8 in. PFA with 3/8 Swage x KF25 adaptor installed tube assembly.
Connect the utility pump power cord to the AC2 UTILITY PUMP socket on the rear of the
control cabinet.
3.3.13 SET UP THE SYSTEM CONTROLLER (PC)
The system controller is a standard personal computer with flat screen display, keyboard and
pointing device. These may be installed at any convenient location that allows the
connection between the PC USB cable from the Control Cabinet and the computer.
Connect the USB cable labeled PC USB that comes out of the back of the Control Cabinet to
a USB port on the back of the system controller (PC).