ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 98
To run the validation test proceed as follows:
Turn OFF the model 3682, open the console door and set the
RUN/CAL/DIAG switch to CAL. Turn the model 3682 back ON and allow
one hour warm up to complete.
Set the model 3682 to <Calibration Manual > mode on the front panel:
Press [A
dvance]
Press [C
alibration Manual]
Press [A
dvance]
Select [S
tandardization]
Select [PS] or [PT]
Press [A
dvance]
Press [A
dvance]
Set the ADCS-601 to absolute measurement mode using the [MODE] button
on the ADCS Tools Main Run Screen.
Vent the ADCS-601 using the [VENT] button on the ADCS Tools Main Run
Screen.
Enter the appropriate <Upper Limit> value on the ADCS Run Screen. Enter
32 inHg when calibrating the Ps transducer and 110 inHg when calibrating
the Pt transducer.
Connect the ADCS-601 TEST(+) port to the 3682 Ps or Pt port.
Select [Run][Run Test Definition] in ADCS Tools.
Select <inHg> as the unit of measure and press [Next]. If the same test was
previously run with the same instrument and pressure range, press [Finish]
to automatically use the previous settings.
In the <Selected DUTs> panel, do NOT select a DUT. If any DUT is
selected, remove it by clicking on it and pressing the [Remove] button.
Press [Next] when ready.