4. OPERATION
Page 99 © 2005 DH Instruments, Inc.
Select the <King 3682 Ps Standardization> or <King 3682 Pt
Standardization> test from the <Select Test Definition> list and press
[Finish]. There are no further initialization steps.
The test completes initialization and starts to run. The status bar updates
during initialization to display the current system operation.
ADCS Tools automatically opens the Data Plot Run Screen (see Section
4.4.3.1) when running a test but there is no relevant data to plot when
running a validation test. To close the Data Plot Run Screen, simply click on
the [x] in the upper right hand corner of the Run Screen.
ADCS Tools sets the test pressure, dwells for 60 seconds, determines the
pressure value set by the ADCS-601 and prompts the user with a dialog box.