ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 100
Enter the ADCS-601 pressure on the model 3682 front panel as the
<Actual> pressure. Then press [ADVANCE] on the model 3682.
Press [OK] when ready.
Repeat step until the last pressure point of the test. Just before the last
point, a special warning dialog box is displayed. The dialog box notifies the
operator that the sequence for the last point is different from the other points.
[OK] must be pressed BEFORE [ADVANCE] is pressed on the model 3682.
On the last test point of the standardization DO NOT press [ADVANCE] on
the model 3682 before [OK] has been pressed on the ADCS Tools dialog box
and the second dialog box indicating it is safe to proceed appears. Pressing
[ADVANCE] on the model 3682 after the last point causes the 3682 to
vent while the ADCS-601 is still at a set pressure which can damage the
ADCS-601.