4. OPERATION
Page 93 © 2005 DH Instruments, Inc.
Select the DUT with <Label> <King 3682> and <Output Label> <Ps> or
<Pt> as desired. Press [Next].
Enter the serial number of the DUT and other identifying information in the
fields provided at the top of the display. Do not change to <IEEE Address>
Also verify that the range is correct, then press [Next].
Select the <King 3682 Ps Validation> or <King 3682 Pt Validation> test
from the list and press [Finish]. There are no further initialization steps. The
test will complete initialization and start to run. The status bar updates
during initialization to display the current system operation.