ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc. Page 94
If the DUT and Test selections are for a <Ps> transducer, a prompt asking
whether to determine the <Offset Check Point> or not is presented. Press
[Yes] on the pop-up window to run this step. It is not necessary to run an offset
Check Point when running a test that will determine as received error or when
trying to evaluate the DUT. In this case press [No] and skip to Step
.
At the next prompt, do NOT press [OK] before having set the model 3682 to
<Calibration Manual> mode. To set <Calibration Manual > mode, on the
front panel of the model 3682:
Press [A
dvance]
Press [C
alibration Manual]
Press [A
dvance]
Select [C
alibration]
Select [PS] or [PT]
Press [A
dvance]
Select [D
ead Weight Tester]
Press [A
dvance]