PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 86
TIME
Read or set the current time of day.
TP
Read the current target pressure.
UCOEF
Convert a pressure in Pascal to pressure in the current units.
UDU
Read or set the user defined pressure unit.
UL
Read or set the upper limit for the current range.
UNIT
Read or set the pressure unit of measure for the current range.
VAC
Read or set the exhaust reference status flag.
VENT
Read, execute or abort a vent process.
VER
Read the PPC3 software version.
ZNATERRn:HI
PPC2+ legacy command. No function in PPC3.
ZNATERRn:LO
PPC2+ legacy command. No function in PPC3.
ZOFFSETn
Read or set the AutoZero P
offset
for the specified Q-RPT.
ZOFFSET:IH
Read or set the AutoZ offset for Hi RPT.
ZOFFSET:IL
Read or set the AutoZ offset for Lo RPT.
ZOFFSETn:HI
Read or set the AutoZ offset for Hi Q-RPT (PPC2+ legacy command).
ZOFFSETn:LO
Read or set the AutoZ offset for Lo Q-RPT (PPC2+ legacy command).
4.4.2 ERROR MESSAGES
Table 17. Error #s and Descriptions
REPLY DESCRIPTION
ERR# 0 “OK”
ERR# 2 “Text argument is too long”
ERR# 3 “Arguments cannot be 0”
ERR# 4 “External device not detected”
ERR# 5 “External device improperly configured”
ERR# 6 “Numeric argument missing or out of range”
ERR# 7 “Missing or improper command argument(s)”
ERR# 8 “External device time-out error”
ERR# 9 “Unknown command”
ERR# 10 “Missing or invalid command suffix”
ERR# 11 “Command missing argument”
ERR# 12 “System overpressured” or “overpressure may result”
ERR# 13 “Text queue overflow”
ERR# 14 “User unit not defined”
ERR# 16 “Generation failure”
ERR# 18 “Command not yet available”
ERR# 19 “Not available with absolute units”
ERR# 20 “Not available with gauge device”
ERR# 21 “User device not defined”
ERR# 22 “Pressure is not stable”
ERR# 23 “Option not available or installed”
ERR# 24 “Unit must be vented”
ERR# 25 “Transducer out of calibration”
ERR# 26 “COM port failed to initialize”
ERR# 27 “Internal device failure”
ERR# 28 “Device failure”
ERR# 29 “Device not available”
ERR# 30 “Must be on range IH”
ERR# 31 “Exceeds upper or lower limit”
ERR# 32 “Not stable enough”
ERR# 37 “Data table is full”
ERR# 38 “Selected range is not available”
ERR# 39 “Data verify error”
ERR# 45 “Argument not allowed”
ERR #46 “Argument cannot be negative”
ERR #52 “Command obsolete”
ERR# 53 “Not Available”