23
The programmes P01 - P11 are pre-programmed EMS programmes for electric
stimulation of the muscle tissue. The programmes U1 - U9 are adjustable EMS
programmes. The respective programme cycles run automatically one after the other and
enhance the effectiveness of the stimulation on the treated muscle or pain region.
(min. = minutes, Hz = vibrations per second, µs = pulse duration in microseconds)
Each programme has an application time of 30 minutes. The corresponding wave forms,
frequencies and pulse durations of the programmes can be found in the programme
overview provided above.
APPLICATION PROGRAMMES
The values shown in the table are factory preset and can be set in the following ranges:
Application time: adjustable from 5 to 90 Minuten
Frequency in Hz: 150 Hz
Pulse duration in
adjustable from 2 to
adjustable fromµs: 50 to 350 µs
Pro-
gram.
Wave form
Applicat.
time
Pulse duration µs
P01
synchronous
30 min.
200 µs
P02
synchronous
30 min.
200 µs
P03
synchronous
30 min.
200 µs
P04
synchronous
30 min.
200 µs
P05
synchronous
30 min.
200 µs
P06
synchronous
30 min.
350 µs
P07
synchronous
30 min.
350 µs
P08
synchronous
30 min.
150 µs
P09
synchronous
30 min.
150 µs
P10
synchronous
30 min.
200 µs
P11
synchronous
30 min.
200 µs
14.0 EMS application programmes
14.1 Pre-programmed EMS application programmes P01 - P11
14.2 Adjustable EMS application programmes U1 - U9
Pro-
gram.
Wave form
Applicat.
time
Pulse duration µs
Factory setting
U1
Changing stimulation
30 min.
200 µs
U2
Changing stimulation
30 min.
200 µs
U3
synchronous
30 min.
200 µs
U4
synchronous
30 min.
250 µs
U5
synchronous
30 min.
350 µs
U6
Changing stimulation
30 min.
350 µs
U7
synchronous
30 min.
350 µs
U8
synchronous
30 min.
150 µs
U9
synchronous
30 min.
250 µs
GB
4. PJN157-14_GA-GB_HHD_TEN250_31.03.14
Montag, 31. M󾾠rz 2014 12:10:42