High Speed (Refrigerated) Micro-Centrifuge
4
WAR NI NG
z This centrifuge is not explosion-proof. Never use explosive or flammable samples.
z Do not install the centrifuge in or near places where inflammable gases are generated
or chemicals are stored.
z Do not place dangerous material within 30cm around the centrifuge.
z Make sure to prepare necessary safety measures before using samples that are toxic,
radioactive or contaminated with pathogenic micro-organisms at your own responsibility.
z If the instrument, rotor and accessories that has been contaminated by solutions with
toxic, radioactive or pathogenic materials, clean it according to the decontamination
procedure that you are specified.
z If you require services at site, please sterilize and decontaminate centrifuge in
advance, and then notice the service center involved in the details of the particular
materials.
z Do not handle the power cord or turn on or off the POWER switch with wet hands to
void electrical shocks.
z For safety purposes, do not enter within 30cm around this centrifuge while it is in
operation.
z While the rotor is rotating, never forcedly release the door lock.
z Unauthorized repairs, disassembly, and other services to the centrifuge except by our
service center are strictly prohibited.
CAUTION
z This centrifuge must be located on one firm and level table.
z Make sure the centrifuge is horizontal before running.
z Make sure the angle between the door and cover is greater than 70 degrees when
open the door.
z Be careful not put your fingers or hands between the door and cover when the door
off.
z Do not move or relocate this centrifuge while it is running.
z When there is fluid in the rotor chamber, please promptly dry with a dry cloth to
avoid sample contamination.