6
2.4 Replacement of motor component
Step 1:
As shown in the left figure,
remove the 3 screws marked by
red circle, remove 3 screws in
the symmetrical side and keep
them well.
Step 2:
Remove the 4 screws marked by
red circles and keep them well,
take down shaker frame.
Step 3:
Take down belt, as shown in the
left figure, remove the 4 screws
marked by red circle, then
replace motor component.