53
Remote Operation
Note:
1. Commands are NOT case sensitive.
2. Every command and response on the Ethernet and serial port should have “\r\n” Carriage
return (0x0D) and Line Feed (0x0A) at the end.
3. Multiple commands with same header can be given in a single command line.
e.g. SYST:IPADDRESS?;TCPPORT?;SERIALNUMBER 2 or ROUTE:SWITCH1
2;SWITCH1?;
note that the commands have to separated by ‘;’
4. The default settings for the Ethernet interface are:
a. IP Address: n.a.
b. IP Port number: 10
c. Gateway address: 200.169.200.5
d. Subnet mask: 255.255.255.0
e. In command SYST:TIMEOUT? The returned value n is in seconds. The default value
is 0. 0 = no timeout.
f. by sending command $/FILE
g. Reboot the matrix
5. In command SYST: SCREENSAVER? and SYST:SCREENSAVER x, the value n and x is in
minutes. The default value is 5 (minutes). 0 = no screen saver. Valid values for x are: 0, 2, 3,
4, 5, 6,….