Preparation
Clean the compressed air supply connection by blowing it clear.
Clean the measuring unit. It must be free of particles and dust.
Check pass-through of the measuring unit (see page 23).
1 Connect the measuring unit to the compressed air supply
connection. The maximum supply pressure is 15 bar. If necessary,
use the adapter.
Outlet pressure 3 bar adjustable at the pressure-reducing valve
with pressure gauge.
Slowly open the compressed air supply valve and flush the system
(allow compressed air to be exhausted through the system).
Flushing time for a regularly maintained system:
2 to 3 minutes.
For other systems:
5 to 6 minutes.
Then close the valve.
Set the timer: 5 minutes.
Strictly follow the specific Instructions for Use.
00321437.e
p
s
1
Oil CO H
2
OCO
2